TOCTAX TOC analyser, 1-100mg/L, 1 channel

OBSOLETE ITEM

TOCTAX TOC analyser, 1-100mg/L, 1 channel
Product #: LPV375.00.00000

OBSOLETE ITEM

Integrated digestion

High level of uptime

Precise measured values due to representative and homogenised sample

Proven self-cleaning system as standard

Low maintenance, simple and rapid

Specifications

Accuracy: ± 4 % + 0.1 mg/L (1.0 - 24.9 - 50 mg/L C)
  ± 5 % + 0.1 mg/L (1 - 100 mg/L C)
Ambient temperature: 5 - 40 °C
Calibration interval: Automatic at selectable intervals (duration: approx. 36 minutes)
Current output: 2 x 0/4 to 20 mA max. 500 Ohm
  PROFIBUS, MODBUS optional
  Service interface RS232
Depth: incl. cooling unit
Dimensions (H x W x D): 1190 mm x 550 mm x 390 mm
Display: Graphics monitor with datalogger and curves display
Enclosure waterproof rating: IP54
Height: incl. cooling unit
Maintenance interval: 3 months
  approx. 1 hour per week
Measuring interval: T 100 16 to 18 min
Measuring principle: Based on DIN EN 1484 – wet chemical peroxide sulphate breakdown process with infrared spectroscopic detection (expulsion method)
Measuring range: 1.00 - 100 mg/L TOC automatic switch over from 1.0-25/50/100mg/l C
Power requirements (Hz): 50 Hz
Power requirements (Voltage): 230 V AC
Process connection (drain): Atmospheric, 4/6 mm connection for waste, 8/11 mm for overflow tray (ID/OD)
Process connection (inlet): 3.2 mm OD
Process connection (installation style): (Analyser) Bypass; homogenisated water sample - wall mounting dry installation, protected against direct sunlight
Process Connection (sample flow): approx. 100 mL/h
Reagent capacitiy: 2 months (standard solution 3-6 months)
Relays: 2 floating limit value contacts 24 V, 1 A
Sample temperature: 5 - 40°C (41 - 104°F)
Special notes: Integrated refrigerator for reagent storage
  The chloride concentration (CI) in the sample must not exceed 400mg/l!
Standard accessories: Set of reagents, standard absorption solution, suitable for 2 month operations
Weight: 43 kg (without reagents)
Width: incl. cooling unit