EVITA OXY Dissolved Oxygen package 2, 0-20 mg/L; 85-264V AC; 50/60 Hz

OBSOLETE ITEM

EVITA OXY Dissolved Oxygen package 2, 0-20 mg/L; 85-264V AC; 50/60 Hz
Product #: 085G4001.52.001

OBSOLETE ITEM

TILTCAL automatic calibration

Easily replaceable sensor - no regeneration!

Self-cleaning design

High accuracy and noise immunity

Connectable directly to PLC

Specifications

Accuracy: ± 0.1 % of FSO at calibrating temperature
  ± 0.2 % of FSO at calibrating temperature
  ± 0.5 % of FSO at 0 - 40 °C
  ± 0.5 % of reading
  ± 0.5 °C of reading
Alarm: On current output (see section “Trouble shooting” of manual)
Altitude: max. 2000 m
Cable length: 10 m 2-wire, screened cable 2 x 0.75 mm2 (18 AWG) cable
Calibration method: Automatic air calibration compensating for temperature, barometric pressure, humidity and salinity by using the TILTCAL or the USC 5000
Communication: HART communication superposed on current output
Current output: 1 x 4 to 20 mA scaleable, 24 V active galvanic isolated, max. load: 800 Ω
  3.80 to 24 mA
Diameter: Transmitter: 240 mm
Diaphragm leakage monitoring: Measurement between anode and a terminal in contact with the measured medium
Display: Back-lit alphanumeric LCD display
EMC: IEC/EN 61000-6-2
  IEC/EN 61000-6-3
Enclosure waterproof rating: IP67 / NEMA 4X
Fuse: Fuse: T500 mA / 250 V (not to be changed by user)
Material: Transmitter: PBT/PC
Measuring principle: Clark cell
Measuring range: 0 - 200 % can be configured
  0 - 20 mg/L can be configured
Number of channels: 1 point-to-point installation
Operating temperature range: 0 - 40 °C (Sensor)
Pollution degree: Class II
Power consumption controller: max. 20 VA
Power consumption transmitter: max. 720 mVA
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 100 - 240 V AC
Relative humidity: 80 % at 31 °C (decreasing linearly to 50% at 40 °C)
Relays: none
Response time: 22 s determined by EVITA OXY 1100
Special notes: Composed of:
USC 5000, EVITA OXY 4100 transmitter (ball), 0-20 mg/l, 10 m cable
EVITA OXY 1100 sensor, 50 µm
mounting bracket for EVITA OXY
Special notes controller: Indoor use
Storage conditions: -40 °C - 70 °C
Strain Reliefs: Pg13.5 (6-12 mm cable diameter)
Time constant: 10 - 300 s (default 40 s)
Weight: 2.7 kg
  2.7 kg