Ultraturb seawater sc, turbidity sensor, 1 m cable

Ultraturb seawater sc, turbidity sensor, 1 m cable
Product #: LPV415.99.82001
Unit Price Contact Us
Ships within 1 week

Large measuring range: 0.0001 - 1.000 NTU

The Hach Ultraturb plus sc Sensor measures turbidity from 0 to 1000 NTU (FNU) for a wide variety of low to medium range turbidity applications. Use it to effectively monitor conventional filtered water processes, effluent water, distribution systems, and elevated tanks where small-sized particulates or air entrapment may be an issue.

Self-Cleaning Sample Chamber Option

The Ultraturb sensor is available with an automatically self-cleaning sample chamber that guarantees stable measured values. The silicon wiper blades inside the sample chamber are held in place with a magnetic coupling wheel to further keep the sample chamber intact.

Ratio Methodology Detects Sample Chamber Fouling

The Ultraturb sensor uses ratio methodology with detection at 90° and 180° to minimize reflections in the measuring cell and on the windows. In this way, fouling of the sample chamber is detected earlier than with a single detection system.

Suitable for seawater, brackish water, raw water and other applications

The new Ultraturb seawater sc is resistant to salt concentrations up to 65 g/L.

What's included

Hach Ultraturb sc, Factory Test Certificate, 4 wipers, Accessory set LZP816, User Manual

Specifications

Accuracy: ± 0.008 FNU ±1% of reading whatever is greater
(0 - 10 FNU)
Automatic wiper: Yes
Cable length: 1 m
Calibration method: Stablcal or dry standard CVM module
Dimensions (H x W x D): 240 mm x 250 mm x 110 mm
Display: Hach SC 200 Digital Controller or Hach

SC 1000 Universal Controller
Enclosure waterproof rating: IP55
Flow rate: 0.2 to 1 L/min
Inlet: 13 mm ID Tubing or fixed connection (PVC system component pipework)
Material: Housing: ASA; Wiper Blade: Silicon; Wiper Shaft: Stainless steel 1.4571
Measurement method: 90° infrared pulse scattered light technique in accordance with DIN EN ISO 7027
Measuring range: 0.0001 - 1000 FNU freely programmable (0.0001 - 250 EBC = 2500 ppm SiO2)
Model: Ultraturb seawater
Operating temperature range: -10 - 40 °C (Display Unit)
Outlet: 13 mm ID Tubing or fixed connection (PVC system component pipework)
Pressure range: 6 bar at 20 °C (87 psi at 68 °F)
Repeatability: ± 0.003 FNU or ± 0.5% of reading (whichever is greater) from 0 - 2 FNU
Response time: 1 - 60 s (programmable)
Sample temperature: Max. 50°C (122°F)
Sensor cleaning: Wiper, automatic
Verification: Stablcal or dry standard CVM module
Warranty: 24 months
Weight: Approx. 1.5 kg
What's included?: Hach Ultraturb sc, Factory Test Certificate, 4 wipers, Accessory set LZP816, User Manual