5500 sc Ammonia Monochloramine Analyser, 1 Channel with External Filter

5500 sc Ammonia Monochloramine Analyser, 1 Channel with External Filter
Product #: 5500.AMC.3.KTO
Unit Price Contact Us

Reliability without all the work

Only Hach can offer a low-maintenance solution with a pressurized reagent delivery system, easy-to-change reagent bottles, a simple troubleshooting menu, and PROGNOSYS predictive diagnostics. You’ll always know if your analyzer is doing its job, so you can do yours.

Easy operation

It has never been easier to monitor your chloramination process online. You’ll be up and running on day one with easy setup, and keep on running with at-a-glance status LEDs, a user friendly interface, color-coded reagent bottles and a simple solution for verifying your results with grab samples.

Improved control of your chloramination process

Continuous online monitoring gives you a more accurate and complete picture of your chloramination process, allowing you to make data-driven decisions. With a dual colorimeter design, you'll get more results faster - keeping you one step ahead of nitrification.

What's included

5500sc Ammonia Monochloramine Analyzer, 1 Channel with installation kit and user manuals, external filter, reagents and standards for one month use

Specifications

Accuracy: ± 5 % or 0.01 ppm (as N) for sample temperature 5 - 40 °C (41 - 104 °F); ±10% or 0.02 ppm for sample temperature 40 - 50 °C (104 - 122 °F), whichever is greater
Alarm: Four; not powered SPDT relays, each rated at 5A resistive, 240 VAC maximum
Certifications: CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000
cETLus (UL 61010-1:2012; NEMA 250:2003; CSA C22.2 No 61010-1:2012)
Australian RCM Marking
Communication: Modbus RS485, Profibus DP (available with external controller), 4-20mA Output
Cycle Time: 4.5 min
Detection limit: 0.01 ppm (as N)
Enclosure waterproof rating: IP56 / NEMA 4X
External Filter: Yes
Flow rate: 100 - 1000 mL/min
Grab sample: Grab Sample In and Grab Sample Out capability
Inlet: 6mm OD push-to-connect fitting for plastic tubing
Light source: Class 1M LED (light emitting diode) with a peak wavelength of 650 nm
Material: Polycarbonate (PC)/ABS case, PC Door, PC hinges and latches, 316 Stainless Steel hardware
Measuring interval: 4.5 minutes
Measuring range: 0.01 - 2 ppm (as N), 0.1 - 10.0 mg/L (as Cl₂)
Mounting: Wall, panel, or table
Number of channels: 1
Number of sample streams: 1 or 2 in programmable sequence
Operating humidity: 5 - 95 % non-condensing
Operating temperature range: 5 - 45 °C (41 - 113 °F)
Options: External Filter
Outlet: 11mm (7/16 in.) ID slip-on fitting
Outputs: Four 0/4-20 mA Outputs, load impedance 600 ohm maximum
Parameter: Total Ammonia, Monochloramine, and Free Ammonia
Pollution degree: 2/II
Power requirements (Hz): 50 - 60 Hz
Power requirements (Voltage): 100 - 240 V AC
Reagent consumption: 1L of reagents every month with 5 minute cycle time
Repeatability: 3 % or 0.01 ppm (as N), whichever is greater
Response time: Within 5 minutes
Sample pressure: 2.5 - 87 psi (0.17 - 6.8 bar)
Sample temperature: 5 - 50 °C (41 - 122 °F)
Storage conditions: -20 - 60 °C (-4 to 140 °F)
Warranty: 24 months
Weight: 20.5 kg (45.19 lb) without reagents and standards, 30 kg (66.14 lb) with reagents, standards and cleaning solution
What's included?: 5500sc Ammonia Monochloramine Analyzer, 1 Channel with installation kit and user manuals, external filter, reagents and standards for one month use