Certifications:
CE Declaration of Conformity conform Machine Directive 2006/42/EC und EMC Directive 2014/30/EU
Dimensions:
(H x W x D) 2.66 m x 2 m x 6 m
Heating system:
Optional, combinable:
- Thermostatically controlled electrical heating (2x 2000 W)
- Thermostatically controlled air conditioning unit with heat pump (3400 W / 4000 W). The air conditioning unit is a wall mounted split version, so no outside ventilation air is blown into the cabinet.
- Thermostatically controlled air conditioning unit with heat pump (3400 W / 4000 W) for corrosive environments with corrosion-resistant coating on the outside heat exchanger unit. The air conditioning unit is a wall mounted split version, so no outside ventilation air is blown into the shelter.
Material:
Shelter material: Sandwich construction made of GRP (glass fiber reinforced polyester) by low pressure Resin Transfer Molding (RTM) using UV resistant gelcoat impregnation process.
Base frame material: SS316L (1.4404), insulated (PU foam), including four SS316L (1.4404) lifting and transport brackets, twistable by 90° mounted to the base frame.
Insulation: 2 cm (0.79'') thickness, K = 1.2 W/m²K
Model:
AS4800
Options:
Complementary equipment:
- Additional lightning protection by aluminium ringline on the roof conform CE regulation mandatory for outside installation
- Additional electrical sockets 230 VAC / 6A (max. 4)
- Sink with drinking water for cleaning hands or equipment
- Hot water boiler
- Oil free compressor
- Insulation: 4 cm (1.57 ''), K = 0.75 W/m²K)
Power:
Each analyser is powered by a black colored service switch with lock position (suitable for LOTO Lock Out Tag Out) and has an individual circuit braker in the local power distribution panel. An electrical power socket is not needed.
According to CE regulations a local main electrical power switch is mandatory. Options (always with lock position, suitable for LOTO Lock Out Tag Out):
- One (1) local main electrical power switch, single phase 230 VAC / 50 Hz - UPS or non-UPS
- Two (2) local main electrical power switches, single phase 230 VAC / 50 Hz - UPS and non-UPS
- One (1) local main electrical power switch, triple phase 3x380 VAC+N / 50 Hz - UPS or non-UPS
- Two (2) local main electrical power switches, triple phase 3x 380 VAC+N / 50 Hz - UPS and non-UPS
- One (1) local main electrical power switch, single phase 230 VAC / 50 Hz - UPS or non-UPS, plus one (1) local main electrical power switch, triple phase 3x380 VAC+N / 50 Hz - UPS or non-UPS
Protection class:
Depending on configurations
Signal junction box:
Optional:
- One (1) combined local signal junction box for analog signals (4 - 20 mA, max. load 500 Ohm) and alarms (potential free contacts, max. 24 V DC). Max. 24 pairs (equals 48 terminals). Mounted on the outside wall of the shelter (shielded by pair).
- Two (2) local signal junction boxes. One box for analog signals (4 - 20 mA, max. load 500 Ohm). One box for alarms (potential free contacts, max. 24 V DC). Max. 12 pairs (equals 24 terminals) per junction box. Mounted on the outside wall of the shelter (shielded by pair).
- Ethernet switch for serial communication, installed in the local power distribution panel. Plus one (1) local signal junction box, mounted on the outside wall of the shelter.
- Ethernet switch for serial communication, installed in the local power distribution panel. Plus two (2) local signal junction boxes, mounted on the outside wall of the shelter.
Ventilation:
Optional:
- Natural ventilation: by means of four ventilation openings installed near diagonal corners on opposite walls at floor and top level.
- Forced ventilation: from outside to inside with a single electrical motor driven fan, which is sized to supply a minimum volume equivalent to eight air changes per hour. Both the fan (ventilation inlet) and the ventilation outlet are installed near diagonal corners on opposite walls at floor and top level.
- Purge ventilation: applicable for corrosive environments. Instead of outside air a purge of instrument air (IA) is used. The purge is distributed by a collector with individual adjustable silencers. With this option there is always a slight overpressure inside the shelter.
Weight:
Approx. 2700 kg (depending on configuration)
What's included?:
AnaShell analytical shelter type AS4800 in ordered configuration; process instrumentation and sample preconditioning pre-assembled and mounted ex works according to order