CA610 Online Fluoride analyser with reagents

CA610  Online Fluoride analyser with reagents
Product #: 5740011
Unit Price Contact Us

Reliable, full-time Fluoride monitoring

The CA610 Fluoride Analyser uses advanced ion-selective electrode (ISE) technology for continuous monitoring of fluoride concentration in drinking water effluent. Accuracy is ensured with precise control of temperature, ionic strength, and pH. Interferences that can bias measurements are virtually eliminated with this electrode.

Patented electrode design with replaceable tip

The proven ISE technology of the electrode is enhanced by an innovative design and manufacturing process. The lanthanum crystal at the tip of the probe is molded into a one-piece, monolithic tip that is threaded for easy removal. This prevents errors and premature failures. Electrodes are equipped with BNC connectors so that they can be independently checked with laboratory electrochemical meters.

Cost effective operation - ultra-low reagent consumption

The electrode of the CA610 analyser is made with a molded fluoride crystal tip. The working life of the electrode is approximately one year. The specially designed tip of the electrode should be replaced every six months. Reagent consumption is very low and maintenance is simple.

Automatic calibration

Two calibration standards are fully enclosed in the CA610 analyser's case. The instrument performs a 2-point calibration at 0.5 and 5.0 mg/L fluoride at user-selected time intervals. Automatic calibration intervals can be set for 1 day to 1 month. Readings and calculations of the most recent calibration is stored in the instrument.

Specifications

Accuracy: ±10% or ±0.10 ppm, whichever is greater
Air purge: 0.1 cfm at 20 psig max.instrument quality air, 1/4-inch OD tubing,
quick-disconnect fitting,
Alarm: 2 SPDT relays (unpowered.), 5A resistive load at 240 Vac. Suitable for sample set point alarms (high or low)
or system warning indicator or system alarm indicator.
Calibration method: Two-point log/linear using 0.5 mg/L and 5.0 mg/L fluoride standards
Compliance certifications: CE approved, ETL listed to UL 1262,ETL certified to CSA 22.2 No. 142,
Cycle Time: 4.2 minutes
Detection limit: 0.10 mg/L
Dimensions (H x W x D): 475 mm x 341 mm x 179 mm
Display: LCD, 3 to 1/2-digit measurement readout and six-character alphanumeric scrolling text line
Drain: 1/2-inch ID flexible hose
Flow rate: 200 to 500 mL/minute
(200 mL/min minimum required)
Inlet pressure: 1 to 10 psig (use sample restriction device
to reduce higher pressure samples)
Measuring range: 0.1 - 10 mg/L fluoride
Mounting: Wall mount
Operating humidity: 90 % at 40 °C max.
Operating temperature range: 5 - 40 °C
Power requirements (Amps): 2.5 A
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 100 - 115/230 V AC
Power supply: selectable, 90 VA max.
Recorder outputs: One isolated recorder output, 4-20 mA (can be adjusted to 0-20 mA); max.
impedance 500 ohm, 130 V isolation
from earth ground.
  Output span is
programmable over any portion of the
measurement range.
Sample temperature: 5 - 40°C (41 - 104°F)
Storage conditions: -40 - 60 °C (without reagents)
Warranty: 24 months
Weight: 11,3 kg ( 25 lb.)