Digital pH Sensor, Stainless

Digital pH Sensor, Stainless
Product #: DPS1.99
Unit Price Contact Us
Available

Exceptional Process pH Sensor Performance with the Differential Electrode pHD Measurement Technique

This field-proven technique uses three electrodes instead of the two normally used in conventional pH sensors. Process and reference electrodes measure the pH differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These process pH sensors provide greater reliability, resulting in less downtime and maintenance.

Lower Maintenance Needs with the Double Junction Salt Bridge

The double junction salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.

Extended Working Life with the Replaceable Salt Bridge/Protector

The unique, replaceable salt bridge holds an extraordinary volume of buffer to extend the working life of the sensor by protecting the reference electrode from harsh process conditions. The salt bridge simply threads onto the end of the sensor if replacement is needed.

Reliability with Built-in Encapsulated Preamp

Encapsulated construction protects the sensor’s built-in preamp from moisture and humidity, ensuring reliable sensor operation. The preamp in the pHD analog sensor produces a strong signal, enabling the sensor to be located up to 1000 m (3280 ft.) from the analyzer.

Patented Technology

The former GLI, now a Hach Company brand, invented the Differential Electrode Technique for pH measurement in 1970. The pHD sensor series takes this field-proven technology to a new level.

Specifications

Accuracy: ± 0.02 pH
Cable connection: Digital
Communication: Modbus
Compliance: Hazardous location, Maritime, CE
Electrode type: General Purpose
Flow rate: 3 m (10 ft.) per second, maximum
Housing material: Stainless Steel
Length: 271.3 mm
Measuring range: -2.0 to 14.0 pH
Mounting: Immersion
Operating temperature range: -5 - 70 °C (23 - 158 °F) pHD and ORP
0 - 50 °C (32 - 122 °F) SS pHD

Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers (matches current recommendation)
Pressure range: Maximum 10.7 bar. 6.9 bar for Digital Sensor at 70°C, and 6.9 bar for Analog Sensor at 105°C.
Repeatability: ± 0.05 pH
Sensitivity: ± 0.01 pH
Sensor cable: 10 m (33 ft.) polyurethane, 4-conductor cable with one shield, rated to 105°C (221°F)
Sensor Thread: 1" NPT
Storage conditions: 4 - 70°C, 0-95% relative humidity (non-condensing)
Temperature accuracy: ± 0.5 °C (± 0.9 °F)
Temperature compensation: Automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature,
additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 °C
Temperature sensor: NTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout
Transmission distance: 1000 m (3280 ft.), maximum when used with a termination box.
  100 m (328 ft.), maximum
Warranty: 24 months
Weight: 0.870 kg
Wetted materials: Stainless steel, PPS, glass, titanium, FKM/FPM o-ring