DR3900 Spectrophotometer with RFID technology

DR3900 Spectrophotometer with RFID technology
Product #: LPV440.99.00001
Unit Price Contact Us
Available

Simple preparation

The boxes and cuvettes are colour-coded for fast and easy parameter and range recognition. Step-by-step illustrated test methods are printed on the box as quick reference or can be called up in the instrument menu.
The DR3900 spectrophotometer uses RFID technology to enable reliable communication with the sample taker and the samples. Simply hold the bottle lable in front of the RFID* sensor. When evaluating the cuvette test, use the tool bar on the touchscreen to assign the result to the sample.

Comprehensive documentation

Measurement results are documented on the detailed level with timestamp, operator ID, absorbance reading, and calculated concentration. The 2D barcode delivers the batch number and expiry date, logged with every result.
For your accreditation the certificate of analysis can be called up just by wiping the reagent box towards the RFID* sensor.

*RFID technology available in all EU countries plus e.g. Great Britain, Norway, Switzerland, Serbia, Macedonia, Turkey, Russia et al. For other countries please ask your local Hach contact person.

Fast execution

A 2D barcode on the LCK cuvettes is automatically read by the Hach DR spectrophotometer to identify the appropriate method and take the measurement.
The vial spins to take 10-fold absorbance readings that will be averaged for result determination to exclude scratches and fingerprints. Instrument calibration verification and high instrument stability all combine to eliminate the need to run reagent blanks.

Customisable

With the ability to store hundreds of user-determined methods, operators are able to tailor the DR3900 to meet the everyday needs of the plant.
Being able to optimise and customise the method portfolio, combined with regular software updates and Claros connectivity, makes the DR3900 the ultimate solution to water quality lab needs.

What's included

Includes: Spectrophotometer DR3900, adapter "A" for 1" round + Accuvac/1 cm rectangular cuvettes, manual in 5 languages (GB, D, F, I, E), power supply 100 - 240V, 47 - 63Hz, Operator RFID Tag.

Specifications

Beam Height: 10 mm
Cuvette compatibility: Rectangular: 10 mm, 50 mm, 1 inch
Round: 13 mm, 1 inch
Data storage: 2000 measured values (Result, Date, Time, Sample ID, Operator ID)
Dimensions (H x W x D): 151 mm x 350 mm x 255 mm
Display: 7" TFT WVGA colour touchscreen
Enclosure waterproof rating: IP30
IBR+: Automatic test recognition, lot control and expiry date check
Interfaces: USB type A (2), USB type B, Ethernet, RFID module
Light source: Gas-filled Tungsten (visible)
LINK2SC: Data exchange with SC1000 controller
Manual languages: German, English, French, Italian, Spanish, Portuguese (PT), Czech, Danish, Dutch, Hungarian, Polish, Romanian, Russian, Slovenian, Swedish, Turkish, Greek, Finnish, Croatian, Bulgarian, Serbian, Slovakian
Max. operating humidity: 80 %
Max. Storage Humidity: 80 %
Operating conditions: 10 - 40 °C
Operating mode: Transmittance (%), Absorbance and Concentration, Scanning
Optical system: Reference beam, spectral
Photometric accuracy: 1 % at 0.50 - 2.0 Abs
  5 mAbs @ 0.0 - 0.5 Abs
Photometric linearity: < 0.5 % - 2 Abs
  ≤ 1 % at >2 Abs with neutral glass at 546 nm
Photometric measuring range: ± 3.0 Abs (wavelength range 340 - 900 nm)
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 110 - 240 V AC
Power Supply: Desk Power Supply
Power supply: External power supply, 100 - 240 V, 50 - 60 Hz
Pre-programmed methods: > 220
Printer compatibility: Supports most office deskjet printers
Quality assurance: Function to schedule and document QA with pass/fail indication
Source lamp: Gas-filled Tungsten (visible)
Specific Technology: RFID for easy method update, sample ID and Certificate of Analysis
Spectral Bandwidth: 5 nm
Standard accessories: None
Storage conditions: -30 °C - 60 °C
Stray light: < 0.1 % T at 340 nm with NaNO2
User interface: Bulgarian, Chinese, Croatian, Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Polish, Portuguese - Brasilian, Portuguese, Russian, Serbian, Slovakian, Slowenian, Spanish, Swedish, Turkish
User programmes: 100
Warranty: 24 months
Wavelength accuracy: ± 1.5 nm (wavelength range 340 - 900 nm)
Wavelength calibration: Automatic
Wavelength range: 320 - 1100 nm
Wavelength reproducibility: ± 0.1 nm
Wavelength resolution: 1 nm
Wavelength selection: Automatic
Weight: 4.2 kg
What's included?: Includes: Spectrophotometer DR3900, adapter "A" for 1" round + Accuvac/1 cm rectangular cuvettes, manual in 5 languages (GB, D, F, I, E), power supply 100 - 240V, 47 - 63Hz, Operator RFID Tag.