Alarm:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient temperature:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analogue outputs:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Calibration:
N.A.
Certifications:
CE compliant / UL certified
Cycle Time:
20 - 30 minutes
40 minutes (+ Total Hardness Ca)
Demineralised water:
For rinsing
Detection limit:
Calcium Hardness: ≤ 10 mg/L
Total Alkalinity: ≤ 20 mg/L
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D):
690 mm x 465 mm x 330 mm
Drain:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flow rate:
100 - 300 mL/min
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences:
Hardness: Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA. Suspended or colloidal organic matter also may interfere with the end point. Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Alkalinity: Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method:
Acid-base titration (alkalinity only); Calmagite EDTA titration with LED dipping probe (combined)
Measuring range:
Ca: 10 - 100 mg/L CaCO₃; Alk: 20 - 500 mg/L CaCO₃
Calcium Hardness: 10 - 100 mg/L CaCO₃; Total Alkalinity: 20 - 500 mg/L CaCO₃
Method:
Titrimetric
Number of sample streams:
1, 2, 4 or 8
Parameter:
Calcium Hardness; Total Alkalinity
Power:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Precision:
Better than 2% full scale range for standard test solutions
Protection class:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Reagent Requirements:
Keep between 10 - 30 °C
Sample pressure:
By external overflow vessel
Sample quality:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample temperature:
10 - 30 °C
Validation:
Automatic; frequency freely programmable
Warranty:
24 months
Weight:
25 kg