Ambient temperature:
5 - 40 °C
Air conditioning and heating options are available.
Communication: digital:
Modbus RTU, Modbus TCP/IP & Profibus (when the Profibus option is selected, the digital output signals are sent through the Profibus converter with its specific communication protocol)
Except for Zone 1 certification then Modbus RTU, Modbus TCP/IP & Modbus TCP/IP Redundant is available
Cycle Time:
From 6.5 minutes, depending on range and application
Data Storage:
Previous 9999 analysis data on screen in the microcontroller memory and storage of data archive for the lifetime of the analyser in the SD/MMC card.
Previous 99 fault data on screen in the microcontroller memory and storage of fault data archive for the lifetime of the analyser in the SD/MMC card.
Detection limit:
0.06 mg/L C
Detection Limit:
B7000i: .06 to 20000 mg C/L
Display:
High contrast 40 character x 16 line backlit LCD with CFL backlight
Drain:
Typically ambient (for applications with high drain pressure, optional systems are available)
EExp / Hazardous Location:
Certification options are available to European Standards, (ATEX Zone 1, Zone 2), North American Standards (Class I Division 2) and IECEx Zone 1
Enclosure waterproof rating:
IP44; optional IP54 with air purge
Exceedance tracking:
Full exceedance tracking to maximum range
Expected Measuring Range:
High range > 5000 mg/L
Humidity:
5 - 85 % (non-condensing)
Languages user interface:
English
Measurement method:
Infrared measurement of CO2 after oxidation
Measuring range:
0 - 20000 mg/L C
Multi-Stream:
1, 2, 4 and 6 streams
Number of channels:
1
Outputs:
1 programmable 4-20 mA analog output signal channel fitted as standard
Oxidation method:
Unique Two-Stage Advanced Oxidation Process using Hyroxyl Radicals
Parameter:
TOC, TIC, TC, VOC, after correlation COD, BOD, TOG
Particle size:
Up to 2 mm, soft particulates
Permissible Chloride range:
Up to 30%
Power requirements (Hz):
50 Hz
Power requirements (Voltage):
230 V AC
Power supply:
TOC/TIC (230VAC)
Range selection:
Automatic or Manual Range Selection
Repeatability:
±3 % of reading or ±0.3 mg/L C, whichever is greater, with Automatic Range Selection
Response time:
TOC 6.5 minutes
Sample inlet temperature:
2 - 60 °C
Sample volume:
Up to 8.0 mL
Service interval:
6 months service intervals
User interface:
Microcontroller with membrane keyboard
Warranty:
12 months
Weight:
90 - 120 kg
Enclosure weight may change depending on system optional features.
What's included?:
Includes: B7000i Analyser, Tubing, Fuses, Ferrules, Drain, Acid & Base Dip Tubes, CO₂ Filter & B7000i User Manual