K1200 LDO sensor for nuclear applications, 0-2000 ppb, 28 mm Orbisphere fitting

K1200 LDO sensor for nuclear applications, 0-2000 ppb, 28 mm Orbisphere fitting
Product #: K1200-S00
Unit Price Contact Us
Ships within 2 weeks

One calibration per year

One zero point calibration per year is all that is needed with the K1200 sensor. Traditional electrochemical (EC) sensors display significant drift after only a few months, demanding regular re-calibration and substantial operator time. Due to its luminescent technology, the K1200 sensor is designed for minimal drift, resulting in it being the most stable sensor with the longest calibration interval in the nuclear industry.

No membranes = two minutes of maintenance

With no membranes to replace and no electrolyte solution to replenish, the K1200 is virtually maintenance-free requiring only two minutes of maintenance per year. Sensor accuracy is unaffected by process changes or low flow events with no polarisation time, eliminating unnecessary operator interventions in radioactive zones. In addition, corrosive or hazardous chemicals are not required, making the annual task faster, easier and safer without reducing measurement precision.

Low cost retrofit

The complete system consists of a Controller, a flow chamber, the K1200 Luminescent Dissolved Oxygen Sensor and a 312xx Hydrogen Sensor for parallel measurement. The sensor is compatible with Hach Orbisphere 28 mm flow chambers, eliminating the need for engineering changes. The installation is fast and easy and does not require special preparation.

A new level of confidence

The K1200 optical sensor is the first to use luminescent measurement technology to measure low level oxygen in nuclear power plants. Since 1978, when the first patent on EC sensors was granted, Hach Orbisphere sensors have set the industry standard for oxygen measurement by delivering confidence to every water chemistry manager. The new K1200 maintains this tradition in the nuclear industry and offers significant operating and cost benefits.

What's included

K1200 Sensor only.
Controller 510, sensor cable, flow-chamber or process connection must be ordered separately.

Specifications

Accuracy: ± 0.8 ppb or 2 % whichever is greater
Ambient temperature: -5 - 50 °C
Analogue outputs: 3 Smart 0/4 to 20 mA (500 Ohms) programmable as linear or tri-linear, configurable to send diagnostics or alarm informations.
Application: Nuclear
Calibration: Single point zero calibration with standard 99.999% nitrogen (quality 5.0) or equivalent oxygen free gas
Certifications: 2004/108/EC - EN 61326-1
Communication capabilities: 3 x 0/4-20 mA; RS485; Ethernet
Detection limit: 0.6 ppb
Display resolution: 0.1 ppb
Flow rate: 50 - 300 mL/min
Measuring range: 0 - 2000 ppb dissolved O2 (DO)
Mounting: 28 mm Orbisphere fitting
Parameter: Oxygen
Power requirements (Voltage): 100 - 230 V AC
Relays: Measurement board: 3 measurement alarm relays (1A-30 VAC or 0.5A-50 VDC), configurable to send diagnostics information.
Main board: 1 system alarm relay (1A-30 VAC or 0.5A-50 VDC).
Repeatability: ± 0.4 ppb or 1 % whichever is greater
Reproducibility: ± 0.8 ppb or 2 % whichever is greater
Response time: (90%) <10 s (gas phase); <30 s (liquid phase)
Sample pressure: 1 - 20 bar absolute (14.5 - 290 psi)
Sample temperature: -5 - 50°C
Sensors: K1200 LDO
Storage conditions: -5 °C - 100 °C
Temperature range: Accurate from -5 - 50 °C
Resistant - temperature from -5 - 100 °C
Warranty: 24 months
Weight: 0.6 kg
What's included?: K1200 Sensor only.
Controller 510, sensor cable, flow-chamber or process connection must be ordered separately.