N-ISE sc Low cost ISE Nitrate probe (immersion) with RFID, 10 m cable

N-ISE sc Low cost ISE Nitrate probe (immersion) with RFID, 10 m cable
Product #: LXV440.99.20001
Unit Price Contact Us
Ships within 1 week

Cost-Effective Trending Information

The N-ISE sc Sensor utilises ion selective electrode (ISE) technology to provide your plant with high level trending information while saving money by eliminating the need for reagents and sample preparation.

Minimal Maintenance with Simple Cartridge Replacement

The sensor works with a Cartrical cartridge that comes factory calibrated, so little maintenance is necessary. Cartridge replacement is simple: unscrew the old cartridge, plug in the new one, and the sensor is ready for measurement. Using RFID* technology, the factory calibration is automatically identified after replacing the cartridge.

Simple, Accurate Calibration

Easy to perform, fail-safe calibration corrections compensate for naturally occurring calibration drift in ISE instruments. An advanced menu structure allows you to perform corrections without manual entry of values via Ethernet, SD card or Bluetooth.

*RFID version available only in US, EU, Canada, Australia, New Zealand, Croatia, Cyprus and Turkey.

What's included

Nitrate probe, sensor cartridge, user manual

Specifications

Accuracy: ± 5% of measured value + 0.2 mg/L NO3-N (with standard solution)
Cable length: 10 m fixed cable at sensor, extendable to 100 m
Calibration method: With Cartrical plus technology: automatic import of factory calibration data from cartridge to probe by RFID*; 1 and 2 point matrix correction
Controller compatibility: SC200, SC1000, SC4500. All controllers sold separately
Data types: current output, relays and bus communication via sc controller
Detection limit: 0.2 mg/L NO3-N
Diameter: 84.5 mm
Flow: < 4 m/s max.
Installation angle: 45 ° ± 15 ° (vertical in flow direction)
Length: 320 mm
Measurement method: Potentiometric ion-selective measurement
Measuring interval: Continuous
Measuring range: 0 - 1000 mg/L NO3-N , Cl-
Measuring uncertainty: ± 5% of the measured value + 0.2 mg/L (with standard solution) NO3-N
Operating temperature range: Air: -20 - 45 °C
Water: 2 - 40 °C
Parameter: NO3-N
pH Range: 5 - 9 pH
Process connection: 1 inch thread
Protection class: IP 68
Repeatability: ±5% (with standard solution) NO3-N
  5% of the measured value
Reproducibility: ±5% of the measured value + 0.2 mg/L NO3-N
Response time: < 3 min
Sample pressure: 0.3 bar max.
Sample temperature: 2 - 40 °C
Sensor cartridge: With Cartrical plus technology: compact housing containing calibrated electrodes for nitrate and chloride, reference system and temperature sensor, all calibrated to each other; typical lifetime 12 months
Storage conditions: 5 - 40 °C sensor cartridge
Warranty: 24 months
Weight: 2.38 kg
Wetted materials: Probe: Stainless Steel 1.4571; ASA + PC, silicon; PVC and PU;
Sensor cartridge: PVC; POM; ABS; Stainless Steel 1.4571; NBR;
What's included?: Nitrate probe, sensor cartridge, user manual