Orbsphere C1100 Ozone Low Pressure Sensor, Stainless Steel, 4 bar

Orbsphere C1100 Ozone Low Pressure Sensor, Stainless Steel, 4 bar
Product #: C1100-LP0
Unit Price Contact Us
Ships within 2 weeks

Maintenance in Less Than Five Minutes

Traditional cleaning processes can take more than thirty minutes to complete, but with the C1100 patented cartridge system, both the set-up and the initial sensor replacement procedure can be performed in less than five minutes without the need for a trained technician.

Low-Touch, Low-Drift Technology

The sensor can remain up to 6 months without calibration. By using a unique platinum guard ring technology, the sensor does not require a separate zero calibration, minimising intervention and low-end drift.

More Accurate than the Lab

The C1100 ozone sensor has a unique air calibration feature that offers operators a quick and easy method to calibrate without requiring an external standard. This air calibration provides better accuracy than the lab and can be done on the operating floor.

Built to Last

The robust stainless steel design makes this sensor extremely robust, with life expectancy going up to 10 years.

What's included

Sensor only.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.

Specifications

Accuracy: ± 0.4 ppb or ±5% whichever is the greater
Application: Municipal
Beverage
Certifications: 2004/108/EC - EN 61326-1
Certified for hazardous environment: No
Detection limit: 0.6 ppb
Flow rate: 350 mL/min (typical)
Material: Stainless Steel
Measuring range: 0 ppb - 50 ppm O₃
  0 ppb - 50 ppm O₃
Membrane: 2957A-LP
Pressure range: Up to 4 bar
Response time: 30 s
Temperature range: Working operating range: -5 - 45 °C
Maximum operating range: -5 - 100 °C
What's included?: Sensor only.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.