SWAS Panel, 2 Parameters

SWAS Panel, 2 Parameters
Product #: LYP105.99.XXXX1
Unit Price Contact Us

Reliable Measurements

A new way of water quality monitoring saves you critical time and effort. Less process downtime due to reliable results delivered by Hach's 9525sc system with Regenerative Cooler and the 5500sc silica analyser.

Complete Solution

Everything from one single source: pre-configured panel, service, digital learning modules, in-house chemical production, and quality control. The single platform minimises time required to teach and learn product operations, getting new systems in use faster.

Easy and Safe Operation

The SWAS panel system is easy to install und requires only low annual maintenance time. Safe operation through shut-off valves for hot samples, safety valves for cooling water, protection screen for hot parts, and certified pressure tests.

Compact Assembly of Panels and Analysers

Sampling and analysers are mounted on a size-optimised single panel. The efficient cooler design ensures the need of less cooling water and a wide sample pressure and temperature range.

What's included

SWAS Panel with DCCP and Silica, including High Pressure Sampling Panel, Hach 9525sc System, Specific Conductivity, Cation Conductivity, Calculated pH and Degassed Cation Conductivity, with Regenerative Cooler (9525.99.20CCR) and 5500sc Silica Analyser, 2 Channel (5500.KTO.S0.A2E); all equipment is mounted on rack panel

Specifications

Flow rate: 200 mL/min for degassed cation conductivity
150 mL/min for 5500sc silica analyser
Plus 350 mL/min for grab sample
Parameter: Degassed cation conductivity, silica
Permissible Chloride range: <250 mg/L for sample temperature 25 - 180 °C
<100 mg/L for sample temperature 180 - 290 °C
<25 mg/L for sample temperature 290 - 550 °C
For higher concentrations Inconel coolers have to be used.
pH Range: 7 - 12 pH
Pressure range: 3 - 6 bar
Sample pressure: <345 bar
Sample temperature: 540 °C maximum
Temperature: Up to 40 °C (pressure drop downstream of cooler: 0.3 - 0.7 bar)
Turbidity: <50 NTU
What's included?: SWAS Panel with DCCP and Silica, including High Pressure Sampling Panel, Hach 9525sc System, Specific Conductivity, Cation Conductivity, Calculated pH and Degassed Cation Conductivity, with Regenerative Cooler (9525.99.20CCR) and 5500sc Silica Analyser, 2 Channel (5500.KTO.S0.A2E); all equipment is mounted on rack panel